| Dimensions of the sterilization chamber (Diameter x H), mm | 213х557 | |||||
| Volume, l | 20 | |||||
| Maximum power of heating elements, kW | 1,4 | |||||
| The sterilizer works from a single-phase alternating current network with voltage and frequency | 220V ±10% and 50 Hz | |||||
| Overall dimensions (WхНxD), mm | 500х400х700 | |||||
| Average service life, years | 10 | |||||
| Weight without spare parts and accessories no more, kg | 45 | |||||
| The emergency shutdown temperature in case of overheating in the chamber is no more than ℃ | 145 | |||||
| Modes of sterilization | ||||||
| 1. Table of programs and cycles of sterilization of packaged medical products made of metals, glass, rubber without using the vacuum drying method | ||||||
| Name | Program 3 | Program 5 | Program 6 | Program 7 | ||
| Sterilization temperature, ℃ | 121±1 | 126±1 | 132±1 | 134±1 | ||
| MPa pressure; kg/cm2 | 0,11(1,1) | 0,14 (1,4) | 0,2 (2,0) | 0,21 (2,1) | ||
| ±0,01 (±0,1) | ±0,01 (±0,1) | ±0,01 (±0,1) | ±0,01 (±0,1) | |||
| Sterilization exposure time, min | 20 (+2) | 10 (+1) | 20 (+1) | 5 (+1) | ||
| Setting the program, monitoring the water in the tank, removing the residual vacuum | ||||||
| Air pumping | ||||||
| Draining condensate from the moisture separator | ||||||
| Heating up to 125℃ | ||||||
| Releasing the pressure until the temperature reaches 105℃ | ||||||
| Heating to the set sterilization temperature | ||||||
| Sterilization time | ||||||
| Pressure reduction in three stages until the temperature reaches 100℃ | ||||||
| Cooling to a temperature of 95℃ | ||||||
| Unloading | ||||||
| 2. Table of programs and cycles of sterilization of empty, porous and packaged medical products using the vacuum drying method | ||||||
| Name | Program 3 | Program 5 | Program 6 | Program 7 | ||
| Sterilization temperature, ℃ | 121±1 | 126±1 | 132±1 | 134±1 | ||
| MPa pressure; kg/cm2 | 0,11(1,1) | 0,14 (1,4) | 0,2 (2,0) | 0,21 (2,1) | ||
| ±0,01 (±0,1) | ±0,01 (±0,1) | ±0,01 (±0,1) | ±0,01 (±0,1) | |||
| Sterilization exposure time, min | 20 (+2) | 10 (+1) | 20 (+1) | 5 (+1) | ||
| Setting the program, monitoring the water in the tank, removing the residual vacuum | ||||||
| Air pumping | ||||||
| Draining condensate from the moisture separator | ||||||
| Heating up to 125℃ | ||||||
| Releasing the pressure until the temperature reaches 105℃ | ||||||
| Heating to the set sterilization temperature | ||||||
| Sterilization time | ||||||
| Depressurization in three stages until the temperature reaches 120℃, 116℃, 106℃ | ||||||
| Vacuum drying | ||||||
| Equalization of pressure in the chamber to atmospheric pressure | ||||||
| Unloading | ||||||
| Delivery set | ||||||
| Sterilizer, unit | 1 | |||||
| Cassette holder, pcs. | 1 | |||||
| Cassette 260x159 pcs. | 4 | |||||
| Cassette 520x159 pcs. | 2 | |||||
| Tank cover, pcs. | 1 | |||||
| Front supports, pcs. | 2 | |||||
| Rear supports, pcs. | 2 | |||||
| Worm clamp 10-15 pcs. | 1 | |||||
| Drainage pipe 10x1 1 m.p., pcs. | 1 | |||||
| Spare parts, 1 pc.: | 10A fuse insert | |||||
| 0,5A fuse insert | ||||||
| Operating instructions, copy | 1 | |||||